Laboratory Devices

KL Cutting Device

KL Cutting Device

  • The test specimen is mounted on the upper plate under a pressure of (9+/-0.2) kPa, and the testing machine is switched off.

  • When the test begins, the specimens are tested according to their wear resistance (number of cycles).

  • A report is prepared based on the changes in the specimens.

KL Heat Testing Device

KL Heat Testing Device

  • The sample is placed on the calorimeter with its outer surface facing upwards.

  • The heat balancer between the heating cylinder and the calorimeter is removed, and the heating cylinder is brought into contact with the sample.

  • A sample is taken during the threshold time, changes are observed, and a report is prepared.

KL Combustion Test Device

KL Combustion Test Device

  • The burning tool is placed under the glove, either vertically along its edge or in the same plane as the middle finger and perpendicular to the glove surface.

  • The distance between the tip of the burning tool and the lower edge of the glove or middle finger is set to 20mm +/- 2mm.

  • The gloves are tested for 3 seconds and 15 seconds for each ignition duration.

  • The product's seams are checked for ignition, and a report is prepared.